FAQ

Home > Servizi > FAQ >
18 Questions found

Why does my hot plate cause an alarm sound?

Everytime the controller detects a defect in the electric circuitry, the entire unit will shut down and the alarm is set on, indicated by a blinking LED light and an alarm sound. The unit’s display will show an error code

Error Code

Meaning

Troubleshooting

E1

One or more plugs of the temperature sensor have been pulled while the stirrer was still running

Connect the temperature sensor back to the main unit. If it doesn’t work, the temperature sensor or its connection cable might be damaged

E2

Temperature Sensor has caused a short-circuit

Replace temperature sensor

E3

Error in internal data storage unit

Contact us!

E4

The installed fan does not operate

Contact us!

If you don't get a satisfactory answer, please click on the "question consulting" message so that we can better solve your problem.

Question

skype
  • Contatti

  • WIGGENS GmbH
  • Wiescher Str. 11a, 42277 Wuppertal
  • Germany
  • Tel.: +49 202 373 29 58-0
  • E-mail: eusupport@wiggens.com

  • WIGGENS CHINA
  • Room 426, Hall A, Office Building M8, No. 1 Jiuxianqiao East Road, Chaoyang District, Beijing
  • Tel.: +86 400-809-2068
  • E-mail: info@wiggens.com

  • www.wiggens.com

AChi siamo Contatti Statement Privacy Policy

Copyright © 2024 WIGGENS CO., LTD. All rights reserved.