FAQ

Home > Servizi > FAQ >
18 Questions found

How do I calibrate the PT100 temperature sensor?

Please follow carefully the following steps in order to calibrate the PT100 temperature sensor:

* Take the calibrators marked as 100 and 140

* Turn on the switch

* Insert the calibrator marked as 100 into the socket of the PT100 sensor

* Press both keys "Up" and "Down" at the same time until "Cal 1" is displayed

* Press the "Start / Stop" key ("100" will be displayed)

* Press the "Start / Stop" key again ("Cal 2" will be displayed)

* Remove the calibrator marked as 100 and insert the calibrator marked as 140

* Press the "Start / Stop" key ("140" and "Off" will be displayed)

* The calibration is completed

If you don't get a satisfactory answer, please click on the "question consulting" message so that we can better solve your problem.

Question

skype
  • Contatti

  • WIGGENS GmbH
  • Wiescher Str. 11a, 42277 Wuppertal
  • Germany
  • Tel.: +49 202 373 29 58-0
  • E-mail: eusupport@wiggens.com

  • WIGGENS CHINA
  • Room 426, Hall A, Office Building M8, No. 1 Jiuxianqiao East Road, Chaoyang District, Beijing
  • Tel.: +86 400-809-2068
  • E-mail: info@wiggens.com

  • www.wiggens.com

AChi siamo Contatti Statement Privacy Policy

Copyright © 2024 WIGGENS CO., LTD. All rights reserved.